Table 1 of
Tu, Mol Vis 2019; 25:391-399.
Table 1. Evaluation of the consistency between the Tonopen and the TonoVet measurement in control healthy monkeys.
Tonometers | IOP ≤21 mmHg (n=90 readings) | |
---|---|---|
Tonopen | TonoVet | |
Mean ± SD | 12.92±2.31 | 12.99±3.04 |
Range | 6-20 | 8-21 |
Δ Mean ± SD | 0.06±2.08 | |
Δ Range | -5-7 | |
95% CI | -4.13 – 4.00 | |
Δ < ±1 mmHg* | 56.67% | |
Δ < ±3 mmHg† | 91.12% | |
P | 0.761 |